APEX formula accuracy
in real clinical practice
The APEX formula was built on a sample drawn from a clinical base of 42,000 surgeries. On held-out eyes that were not used to develop the formula, APEX matches Kane in accuracy and outperforms the classic formulas.
in the APEX clinical base
for the table leader
among the compared formulas
Formula comparison
Computation was performed on 960 held-out eyes that were not used to develop the formula. Each formula was computed with its own optimized constant under equal conditions. Kane was included as an external benchmark. Sorted by the FPI index.
| Formula | MAE (D) | MedAE | ±0.5 D | FPI | vs APEX |
|---|---|---|---|---|---|
| 1APEX | 0.348 | 0.278 | 76% | 0.635 | leader |
| 2Kaneexternal | 0.352 | 0.288 | 75% | 0.623 | comparable |
| 3PEARL-DGS | 0.370 | 0.296 | 75% | 0.604 | lower |
| 4Castrop | 0.408 | 0.332 | 69% | 0.541 | lower |
| 5Holladay 1 | 0.402 | 0.322 | 70% | 0.537 | lower |
| 6T2 | 0.415 | 0.335 | 68% | 0.531 | lower |
| 7Hoffer Q | 0.423 | 0.345 | 68% | 0.522 | lower |
| 8Haigis | 0.399 | 0.327 | 69% | 0.522 | lower |
| 9SRK/T | 0.427 | 0.347 | 67% | 0.508 | lower |
| 10MIKOF | 0.427 | 0.348 | 68% | 0.507 | lower |